• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER HEMODIALYTSIS NON IMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER HEMODIALYTSIS NON IMP Back to Search Results
Catalog Number CS-22122-F
Device Problems Difficult to Insert (1316); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reports: the middle of the guidewire was bent and couldn't get in.The md took out another new product and finished the procedure.
 
Manufacturer Narrative
(b)(4).The customer returned an opened hemodialysis kit containing a spring wire guide (swg) within its advancer tubing, catheter, dilator, ars, introducer needle, and various other components for evaluation.Visual inspection of the swg revealed one kink in its body.Microscopic examination of the swg confirmed the kink and that both welds were in place.No other defects or anomalies were observed in the other components.The kink in the swg body was measured at 453 mm from the proximal weld.The total length of the swg measured 601 mm, which is within specifications of 596-604 mm per swg graphic.The outer diameter of the swg measured 0.854 mm which is within specifications of 0.838-0.877 mm per swg graphic.The swg was inserted into the returned ars, 18 ga introducer needle, catheter, and dilator to functionally test.The swg only encountered resistance at the kink for all four components.A manual tug test confirmed that both the distal and proximal welds were intact.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The instructions-for-use (ifu) provided with this kit warns the user, "do not apply excessive force in removing guide wire or catheters.If withdrawal cannot be easily accomplished, a chest x-ray should be obtained and further consultation requested." the report that the guide wire kinked during use was confirmed through examination of the returned sample.The guide wire contained one kink in its body.The returned guide wire met all relevant dimensional requirements and a device history record review based on sales history did not identify any manufacturing related issues.Based on the condition of the guide wire and the report that the damage was observed during use, unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
The customer reports: the middle of the guidewire was bent and couldn't get in.The md took out another new product and finished the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM
Type of Device
CATHETER HEMODIALYTSIS NON IMP
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10001151
MDR Text Key189053568
Report Number3006425876-2020-00354
Device Sequence Number1
Product Code MPB
Combination Product (y/n)N
PMA/PMN Number
K993933
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCS-22122-F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2020
Date Manufacturer Received05/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-