ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Capture (1081); Difficult to Insert (1316); Device Dislodged or Dislocated (2923)
Patient Problems Chest Pain (1776); Dyspnea (1816)
Event Date 04/03/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient presented in hospital due to occasional chest distress and shortness of breath.Upon interrogation, loss of capture and lead dislodgement were observed on the left ventricular lead.During lead repositioning procedure, the guide wire could not pass through the lead.The lead was then explanted and replaced.The patient was stable.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10006166
MDR Text Key189210964
Report Number2938836-2020-02890
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000067970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2020
Was the Report Sent to FDA? No
Date Manufacturer Received06/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight102
-
-