Device evaluation by manufacturer: a field service engineering (fse) was at customer's site to resolve reported event.Fse confirmed the error by reviewing the error log and reproduced error by attempting a sample run.While troubleshooting, fse noticed the rack pusher foot was out of position.Fse adjusted the home sensor flag for the x1 motor, then ran rack rotation without errors.Fse checked rack presence sensors and all was operating as expected.Fse successfully completed a daily check run without errors.Customer performed a quality control run, which passed without errors and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number 11358105.There were no other similar complaints found during the searched period.The aia-900 operator's manual under section 12 - flags and error messages states: [2300] s.Loader step feed failure cause: the pitch sensor s071 failed to go on after the step feed.Action: check to see if there is any impediment to the sample rack feed.If the trouble reoccurs, contact the tosoh local representatives.Check s071 and the step feed mechanism.The most probable cause of the reported event is due to misaligned x1 sensor flag.
|