Model Number US906
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2018
Event Type  malfunction  
Manufacturer Narrative
Note: a retrospective review of potential serious injury complaints was performed.This mdr was identified as a reportable malfunction as a result and filed as part of the review activities.The complained product was not returned and the batch # was not provided.Due to the time frame, it was reasonable to assume that the product fell within the scope.
 
Event Description
It was reported there was an issue with the orange locks.The reporter indicated that several us906 do not lock as they should then breaking as they try to lock them in place.Also, the color change is inconsistent.For example, on a given container, they will change from blue to brown on one side of the container, but on the other side the color will not change at all.The locks are retried until one works.No serious injury or death.No delay in surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM STER LOCKS ORANGE
Type of Device
STERILE TECHNOLOGY
Manufacturer (Section D)
SPS MEDICAL (CROSSTEX)
6789 w. henrietta road
rush, ny
Manufacturer (Section G)
AESCULAP INC
3773 corporate parkway
center valley, pa
Manufacturer Contact
lindsay chromiak
3773 corporate parkway
center valley, pa 
2581946507
MDR Report Key10009668
MDR Text Key189347835
Report Number2916714-2020-00122
Device Sequence Number1
Product Code KCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K890759
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUS906
Device Catalogue NumberUS906
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial