• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC STEAM STER LOCKS ORANGE; STERILE TECHNOLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC STEAM STER LOCKS ORANGE; STERILE TECHNOLOGY Back to Search Results
Model Number US906
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2016
Event Type  malfunction  
Manufacturer Narrative
A retrospective review of potential serious injury complaints was performed.This mdr was identified and filed as part of the review activities.Supplier: (b)(4).Actions taken: capa opened by aesculap inc.Corrective action: sps- in october 2016, a labeling change had been made to state that this product should be kept out of direct sunlight, away from heat sources, away from hydrogen peroxide, and other sterilants.Description: there were an increased number of complaints on the indicator dot labels reverting back to blue or not fully turning.The supplier was made aware of the complaints and various lots were investigated.During the investigation, ink manufacturing process was reviewed and verified.No changes, other than going to a lead-free ink were made in the process or to the raw materials associated with making this product.Additionally, a comprehensive review was conducted with the ink manufacturer on the chemical makeup of the ink.No changes were made or found.Lastly, multiple tests and studies were conducted in an attempt to determine the root cause of the issues being reported by customers.Conclusion: at this time the issues of reversion and inconsistent color change were able to be duplicated.Through testing it was found that both lead and lead-free indicator labels would be adversely affected by being exposed to hydrogen peroxide post-sterilization.In all test cases, a reversion would occur once the processed label was exposed to hydrogen peroxide.The lead-free indicator labels were adversely affected by the exposure to direct sunlight and hydrogen peroxide.In hydrogen peroxide exposure prior to sterilization, an inconsistent color change was experienced and led to a reversion over the course of testing.Additionally, exposure to direct sunlight has caused the indicator label to revert and fluorescent light has begun to fade these indicator labels.Lead indicator labels were not as sensitive to these exposure types but did show signs that these exposures affected the product.During pre-sterilization hydrogen peroxide exposure, it was found that the lead indicator labels displayed an inconsistent color change.Both direct sunlight and fluorescent light did not appear to adversely affect the product.Corrective action: see above.
 
Event Description
It was reported that there was an issue with the orange locks.A few days after sterilization of the container, the indicator dot changes back to blue.It appears as though the set never went through the sterilization process.The malfunction caused a surgical delay for an unspecified procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM STER LOCKS ORANGE
Type of Device
STERILE TECHNOLOGY
Manufacturer (Section D)
AESCULAP INC
3773 corporate parkway
center valley, pa
Manufacturer (Section G)
AESCULAP INC
3773 corporate parkway
center valley, pa
Manufacturer Contact
lindsay chromiak
3773 corporate parkway
center valley, pa 
MDR Report Key10014124
MDR Text Key189490183
Report Number2916714-2020-00155
Device Sequence Number1
Product Code KCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUS906
Device Catalogue NumberUS906
Device Lot Number1601
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2016
Initial Date FDA Received04/30/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-