• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP, INC. STEAM STER LOCKS ORANGE; STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP, INC. STEAM STER LOCKS ORANGE; STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES Back to Search Results
Model Number US906
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2018
Event Type  malfunction  
Manufacturer Narrative
Note: a retrospective review of potential serious injury complaints was performed.This mdr was identified as a reportable malfunction as a result and filed as part of the review activities.
 
Event Description
It was reported there was an issue with the orange locks.The reporter indicated that several us906 do not lock as they should then breaking as they try to lock them in place.Also, the color change is inconsistent.For example, on a given container, they will change from blue to brown on one side of the container, but on the other side the color will not change at all.The us906 locks are retried until one works.No serious injury or death.No delay in surgery.
 
Manufacturer Narrative
Additional information - d4 (udi).Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM STER LOCKS ORANGE
Type of Device
STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES
Manufacturer (Section D)
AESCULAP, INC.
3773 corporate parkway
center valley PA 18034
MDR Report Key10015914
MDR Text Key189577056
Report Number2916714-2020-00157
Device Sequence Number1
Product Code KCT
Combination Product (y/n)N
PMA/PMN Number
K890759
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUS906
Device Catalogue NumberUS906
Device Lot Number583835-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/16/2018
Initial Date FDA Received04/30/2020
Supplement Dates Manufacturer Received10/07/2020
Supplement Dates FDA Received11/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-