MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX¿ SELF-ADHERING MALE EXTERNAL CATHETER; MEC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX¿ SELF-ADHERING MALE EXTERNAL CATHETER; MEC Back to Search Results
Model Number 33301
Device Problems Nonstandard Device (1420); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the catheters were painful to remove from the patient and should come with instructions to trim the area.
 
Manufacturer Narrative
The reported event could not be confirmed as no sample was returned for evaluation.A potential failure mode could be ¿too high viscosity¿ with a potential root cause of ¿viscometer failure or mechanical failure¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the catheters were painful to remove from the patient and should come with instructions to trim the area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX¿ SELF-ADHERING MALE EXTERNAL CATHETER
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10018221
MDR Text Key189985346
Report Number1018233-2020-02970
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741070730
UDI-Public(01)00801741070730
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Type of Report Initial,Followup
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number33301
Device Catalogue Number33301
Was Device Available for Evaluation? No
Date Manufacturer Received05/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-