Model Number G247
Device Problems Inaccurate Synchronization (1609); Unexpected Therapeutic Results (1631)
Patient Problem Ventricular Fibrillation (2130)
Event Date 01/30/2020
Event Type  Injury  
Manufacturer Narrative
Records indicate this device remains in service.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported this device recorded an appropriate ventricular episode due to the ventricular rate being greater than the atrial rate.The ventricular rate started around 190 beats per minute (bpm) with anti-tachycardia pacing (atp) delivered.The atp did not covert the rhythm and a 5 joule shock was delivered.This shock accelerated the arrhythmia to and the ventricular rate was 335 bpm.A 41 joule shock was delivered which did not convert the rhythm.A second 41 joule shock was delivered which was successful at converting the rhythm.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key10021377
MDR Text Key189757137
Report Number2124215-2020-09894
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/27/2019
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number172311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2020
Initial Date FDA Received05/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age82 YR
-
-