• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC LEVEL 1 HOTLINE LOW FLOW SYSTEMS; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC LEVEL 1 HOTLINE LOW FLOW SYSTEMS; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the fluid warmer had broken lcd.The product problem was discovered during preventative maintenance.The device was fully operational with damage to only the display.There was no patient involvement.
 
Manufacturer Narrative
Evaluation results: one level 1 hotline low flow system was returned for investigation in used condition.Visual inspection revealed that the enclosure was heavily stained and marked, both covers were cracked and marked, the line cord was worn, and the pole clamp handle was broken.The investigator subsequently filled the tank with water, attached the temperature check, plugged in the line cord, and turned on the power switch.The customer reported product problem (broken lcd) was confirmed during testing.The product problem occurred because of a damaged pcb board.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.The problem source of the reported product problem was unknown.A root cause was not established.The damaged pcb and aforementioned components were replaced.Preventative maintenance was then performed.The device subsequently passed functional testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC
6000 nathan lane north
minneapolis, mn
MDR Report Key10026289
MDR Text Key189915693
Report Number3012307300-2020-03652
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2020
Initial Date Manufacturer Received 04/06/2020
Initial Date FDA Received05/05/2020
Supplement Dates Manufacturer Received05/08/2020
Supplement Dates FDA Received06/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-