A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed the error on the error log and noted the bf table was replaced recently.During the evaluation, the fse discovered the alignment of the x-axis cup transfer to the bf table had shifted.The fse reset the alignment and performed a system test.Daily check, calibration, and quality control (qc) results were within the acceptable range.The aia-2000 analyzer returned to normal operation.No further field action required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 06mar2019 to aware date 06apr2020.No other similar complaints were identified during the search period.The aia-2000 operator's manual states the following: [4163] x-axis cup transfer z-axis home overrun.Cause: the home sensor activated improperly after movement of the x-axis cup transfer z-axis.If this occurs, the current measurement result will be flagged (mf flag) and new measurements will be suspended.Solution: contact tosoh service center or local representatives.The probable cause of the issue is attributed to the misalignment of the x-axis cup transfer to the bf table.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|
The customer reported error 4163 message after approximately fifty (50) samples on the aia-2000 analyzer.The customer noted the cup missed the incubator.A version up and cleaning had previously been performed, but the error continued.A field service engineer (fse) was dispatched to address the reported issue, which resulted in a delay reporting of follicle-stimulating hormone (fsh) test results.There was no report of patient intervention or adverse health consequences due to the event.
|