The device was received for evaluation.Evaluation of the device determined that the gas tubes were dirty inside.A strange liquid was found inside the gas tubes which invaded into the 1st regulator unit, electropneumatic proportional valve , main fold pipe and the k-connector unit.In addition, the front panel was found broken at the bottom side.The device was placed for repair.As stated on the ifu and as a preventive measure, the user manual states : keep fluids away from all electrical equipment.If fluids are spilled on or into the high flow insufflation unit, stop operation of the unit immediately and contact olympus.Always use the uhi-4 as outlined in this instruction manual.Improper use will not only impair functions and prevent optimum performance, but may cause equipment damage and/or complications.Before each use, always inspect the equipment as outlined in this instruction manual.
|
The following fields were updated: g4, g7, h2, h6, h10.This supplemental report is being submit to provide the results of the manufacturer's investigation.A review of the device history record (dhr) was conducted as part of the investigation.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met specifications upon release.The root cause was not identified.Probable causes include damage from user handling.Liquid had entered the gas tube and this is assumed to be the cause of the regulator damage.Olympus will continue to monitor the field performance of this device.
|