TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number 1002717
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.Per the user guide: always twist the luer-lock connection between the cartridge tubing and the infusion set tubing an extra quarter of a turn to ensure a secure connection.A loose connection can cause insulin to leak, resulting in under delivery of insulin.This can cause high blood glucose.
 
Event Description
It was reported that the cartridge and infusion set luer lock became disconnected.Customer's blood glucose (bg) was 486 mg/dl and manual insulin injection was administered to address bg.Customer reconnected luer lock to resolve the issue and insulin delivery was resumed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key10036351
MDR Text Key190318227
Report Number3013756811-2020-44152
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1002717
Device Catalogue Number1002684
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 04/16/2020
Initial Date FDA Received05/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age22 YR
-
-