• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SYRINGE; SALINE VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306594
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/25/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: based on no sample, the investigation concluded, bd was not able to verify the indicated failure.This is the 1st complaint for lot # 9196132 for this type of defect or symptom.Root cause description: no root cause can be determined as no samples were received.Rationale: capa not required at this time.
 
Event Description
It was reported that the bd posiflush¿ syringe leaked while flushing the indwelling needle for a patient with "mild acute pancreatitis and hypertension".The following information was provided by the initial reporter, translated from (b)(6) to english: "due to mild acute pancreatitis and hypertension, the patient received a doctor's appointment.On (b)(6) 2020, the medical staff used the prefilled syringe to flush the disposable intravenous indwelling needle, and found the liquid extravasation of the irrigator.He stopped using the irrigator immediately, and replaced the new irrigator, and sealed the catheter successfully.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key10036637
MDR Text Key195258686
Report Number1911916-2020-00447
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number306594
Device Lot Number9196132
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-