• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/09/2020
Event Type  malfunction  
Manufacturer Narrative
The country of origin is (b)(6).
 
Event Description
The initial reporter received a questionable elecsys ft4 iii assay result from the cobas 8000 e 801 module serial number (b)(4).For patient results.The questionable result was reported outside the laboratory.
 
Manufacturer Narrative
Based on the available information, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10039121
MDR Text Key199494378
Report Number1823260-2020-01199
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 05/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot Number432844
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/22/2020
Initial Date FDA Received05/08/2020
Supplement Dates Manufacturer Received04/22/2020
Supplement Dates FDA Received05/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-