The following fields were updated due to additional information: device available for eval yes, returned to manufacturer on: 2020-05-21.Investigation summary: a device history record review was performed for provided lot number 9248141 and the review did not reveal any detected quality issues during the production process that could have contributed to this reported incident.To aid in the investigation of this issue, both picture and physical samples were provided for evaluation by our quality team.Through examination of the samples, the presence of foreign matter/brown stains on the product packaging was confirmed.It has been determined that the staining is a result of the steam sterilization process.The integrity of the product and the sterile barriers have not been affected.These spots appear only on the outside of the packaging and do not permeate in any way onto the product.Furthermore, microbial permeability testing, cytotoxicity testing, and testing for residual solvents and volatile species has been completed on the packages displaying the brownish stain.The testing confirmed that they do not present any risk to the use of the product and have no impact on the effectiveness, sterility, quality or safety of bd posiflush¿ xs 10ml saline flush syringe.We are currently exploring several options to reduce and eliminate any brown staining that occurs on the packaging.
|