Catalog Number 306572
Device Problem Packaging Problem (3007)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2020
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd posiflush¿ xs pre-filled flush syringe nacl 0.9% had foreign matter inside.This was discovered before use.The following information was provided by the initial reporter: new, dirty bd saline xs syringes.
 
Event Description
It was reported that bd posiflush¿ xs pre-filled flush syringe nacl 0.9% had foreign matter inside.This was discovered before use.The following information was provided by the initial reporter: new, dirty bd saline xs syringes.
 
Manufacturer Narrative
The following fields were updated due to additional information: device available for eval yes, returned to manufacturer on: 2020-05-21.Investigation summary: a device history record review was performed for provided lot number 9248141 and the review did not reveal any detected quality issues during the production process that could have contributed to this reported incident.To aid in the investigation of this issue, both picture and physical samples were provided for evaluation by our quality team.Through examination of the samples, the presence of foreign matter/brown stains on the product packaging was confirmed.It has been determined that the staining is a result of the steam sterilization process.The integrity of the product and the sterile barriers have not been affected.These spots appear only on the outside of the packaging and do not permeate in any way onto the product.Furthermore, microbial permeability testing, cytotoxicity testing, and testing for residual solvents and volatile species has been completed on the packages displaying the brownish stain.The testing confirmed that they do not present any risk to the use of the product and have no impact on the effectiveness, sterility, quality or safety of bd posiflush¿ xs 10ml saline flush syringe.We are currently exploring several options to reduce and eliminate any brown staining that occurs on the packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE NACL 0.9%
Type of Device
NA
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
MDR Report Key10045746
MDR Text Key204181507
Report Number9616657-2020-00073
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 05/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2022
Device Catalogue Number306572
Device Lot Number9248141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2020
Initial Date Manufacturer Received 04/20/2020
Initial Date FDA Received05/11/2020
Supplement Dates Manufacturer Received04/20/2020
Supplement Dates FDA Received05/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-