This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d10, g4, g7, h2, h3, h4, h6, h8, h10.Product review of the zimmer air dermatome serial number (b)(6) by flextronics on 25 may 2020 revealed that the unit was out of calibration and the control bar position was not correct.Repair of the device was performed by a flextronics on 25 may 2020 which included replacement of the following: control bar: pn 06001810352, ln 64370635.The device, serial number (b)(6), was then tested and functioned properly.It was repaired, inspected and tested.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|