Model Number 22-4038
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2020
Event Type  Injury  
Event Description
It was reported that during an arthroscopy the first-pass suture was not passing through the rotator cuff.The needle was bent.The procedure was completed using an smith & nephew accupass device.A delay of 10 minutes was reported.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3, h6: the disposable firstpass suture passer, self device, used in treatment, was returned for evaluation.A relationship between the device and reported incident was not established.A review of manufacturing records for the reported lot number 2040498 found no non-conformance's or anomalies during manufacturing process related to the reported event.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.Review of the product instructions for use found adequate warnings and precautions to prevent damage to the device during use.Risk management documents were reviewed finding no additional risks that require to be added to the reference document.Visual inspection shows no manufacturing abnormalities on the device.The instrument was returned with a closed top bracket.The needle is bent.The first step of the trigger stayed locked in the closed position and cannot be released.An attempt was made to test the basic functions of the instrument but due to the damage it is not possible.The complaint was verified.An exact root cause cannot be determined with confidence; however, potential factors unrelated to the design or manufacture of the device that may lead to the failure reported include, but are not limited to: (1) excessive force (2) tissue thickness (3) debris or damage on the tip between passes.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE FIRSTPASS SUTURE PASSER, SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10050159
MDR Text Key190757271
Report Number3006524618-2020-00253
Device Sequence Number1
Product Code HWQ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/17/2022
Device Model Number22-4038
Device Lot Number2040498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2020
Date Manufacturer Received06/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-