• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS AUTOLOG WASH KIT; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS AUTOLOG WASH KIT; APPARATUS, AUTOTRANSFUSION Back to Search Results
Model Number ATL2001
Device Problem Crack (1135)
Patient Problem Blood Loss (2597)
Event Date 04/27/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use, the customer reported that during the autotransfusion operation, the patient's blood entered the rotating centrifugal bowl of the autolog washkit and red blood cells were collected.The bowl then ruptured (cracked) due to the pressure in the centrifugal bowl.The product was replaced.There was blood loss as a result of the reported issue and a blood transfusion was required.There was no damage reported to any of the device packaging.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOLOG WASH KIT
Type of Device
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
MDR Report Key10050940
MDR Text Key190762088
Report Number2184009-2020-00024
Device Sequence Number1
Product Code CAC
UDI-Device Identifier00613994520456
UDI-Public00613994520456
Combination Product (y/n)N
PMA/PMN Number
K093535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/05/2021
Device Model NumberATL2001
Device Catalogue NumberATL2001
Device Lot Number218332385
Was Device Available for Evaluation? No
Date Manufacturer Received06/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
Patient Weight81
-
-