SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION EXL WITH LM Back to Search Results
Model Number DIMENSION EXL WITH LM
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/22/2020
Event Type  malfunction  
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc) and a siemens customer service engineer (cse) was dispatched to the customer's site and inspected the instrument.Cse adjusted the top seal solenoid and verified that there was proper operation with no seal errors.System was operational upon departure.Cse confirmed that cause of the burst top cuvette seal was the customer squeezing the cuvette.The instrument is performing according to specifications.No further evaluation of this device is required.
 
Event Description
Customer was splashed with biohazardous waste while emptying an unsealed film cuvette on their dimension exl with lm instrument.The operator was wearing personal protective equipment (ppe) when emptying the waste and no treatment was sought by the operator.There were no slips, trips or falls related to the splash.There are no known reports of patient intervention or adverse health consequences due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION EXL WITH LM
Type of Device
DIMENSION EXL WITH LM
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark, de
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
101 silvermine road
registration number: 1226181
brookfield, ct
Manufacturer Contact
eric so
511 benedict ave
tarrytown, ny 
5243067
MDR Report Key10054474
MDR Text Key195648250
Report Number2517506-2020-00143
Device Sequence Number1
Product Code JJE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130276
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION EXL WITH LM
Device Catalogue Number10486890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/22/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse