• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC. FENESTRATED BIPOLAR FORCEPS; SYSTEM, SURGICAL, COMPUTER CONTROLLED INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC. FENESTRATED BIPOLAR FORCEPS; SYSTEM, SURGICAL, COMPUTER CONTROLLED INSTRUMENT Back to Search Results
Model Number 470205
Device Problem Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/11/2020
Event Type  Injury  
Event Description
Fenestrated bipolar malfunction.After multiple attempts of troubleshooting bipolar the surgeon noticed a single piece of cable sticking out of the jaws of the instrument.The instrument would not release after several attempts of troubleshooting this new issue.Eventually we learned the metal portion of the instrument was lifting away from the shaft alerting the davinci of an obstruction.The surgeon was able to safely remove the instrument and its entirety along with the metal trocar.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FENESTRATED BIPOLAR FORCEPS
Type of Device
SYSTEM, SURGICAL, COMPUTER CONTROLLED INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL, INC.
MDR Report Key10055852
MDR Text Key191115189
Report NumberMW5094492
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874112359
UDI-Public(01)00886874112359
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470205
Device Catalogue Number470205
Device Lot NumberN11190806 0061
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/13/2020
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR
-
-