One smiths medical level 1 hotline low flow system was returned for analysis with a wear and tear damaged enclosure and faulty pump assembly.During analysis, the reported issue was able to be duplicated.It was noted that the pump assembly was faulty and didn't pump water as it should and was the cause of the reported issue.Service replaced the pump assembly to correct the reported issue.Based on the evidence, the complaint was confirmed, and the problem source of the reported event was noted to be faulty pump.
|