The reported field event of a reset was confirmed in the lab.The device went to backup vvi due to a power-on-reset (por).Interrogation of the device revealed the device was above the elective replacement indicator (eri) when received.The device was tested on the bench: telemetry, pacing, sensing, impedance, hv output, hv shock and patient notifier were tested on the bench.No anomaly was detected.The reset could not be reproduced in the lab; hence the root cause of the anomaly remains undetermined.
|