SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Mechanical Problem (1384)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2020
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that patient's right driveline could not detach due to a displaced cpc connector spring.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup freedom driver.
 
Manufacturer Narrative
The right cpc connector from freedom driver s/n (b)(6) driveline was returned to syncardia for evaluation.Visual inspection confirmed the customer-reported issue of a displaced spring.This failure mode aligns with the failure mode determined in capa- 0266, cpc connector issues.Syncardia has a corrective and preventive action (capa) to address the issue of missing or displaced springs in cpc connectors.From the results of the capa investigation, the primary root cause of spring displacement within the cpc connectors is the use of a wire tie within the cpc connector body (which can come in contact with the spring).Clinicians and patients are trained to thread a wire tie beneath the thumb release tab of the cpc connectors to prevent accidental disconnection of the cannulae from the drivelines.It is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch out.Syncardia capa-0266, cpc connectors, is currently at step 5 action plan.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10062400
MDR Text Key204580234
Report Number3003761017-2020-00129
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2020
Initial Date Manufacturer Received 04/21/2020
Initial Date FDA Received05/15/2020
Supplement Dates Manufacturer Received04/21/2020
Supplement Dates FDA Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
-
-