Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of "(sn: (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message" was confirmed during the functional testing and based on the archive data review.The root cause for the occurrence of ua07 error was due to defective load cells as a result of damage sustained by user mishandling and/or aging of the device.The autopulse platform is a reusable device and was manufactured in march 2010 and is ten years old, beyond its expected service life of 5 years.Visual inspection of the returned platform was performed, and no issues were observed.Review of the archive data showed multiple user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error messages; thus, confirming the reported complaint.During functional testing, user advisory (ua) 07 error message displayed upon power up the device; thus, confirming the reported complaint.In addition, the load characterization check was performed and verified that both load cells were out of specification.The load cells were replaced to address the user advisory (ua) 07 error.Following the service, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The brake gap inspection was performed and verified the brake gap was within the specification.Load cell characterization test was performed and confirmed both cell modules are functioning within the specification.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
The autopulse platform (sn: (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message.Patient use information was requested, but no additional information was provided; therefore, patient use is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10063915
MDR Text Key191477890
Report Number3010617000-2020-00520
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000130
UDI-Public00849111000130
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-