• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).(b)(4).
 
Event Description
The initial reporter received questionable high elecsys ft4 iii assay results for one patient from a cobas 8000 e 801 module serial number (b)(4).The customer reported out the results to a physician who asked for re-measurement of the sample.Sample from the patient was submitted for investigation and was tested on an architect analyzer and an e 801 module.Refer to the attachment on the medwatch for all patient data.
 
Manufacturer Narrative
Calibration and qc information was requested but not provided.Based on the available data, a general reagent issue can be excluded.The observed differences in ft4 values generated with the roche assay and the abbott architect assay are most likely caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standardization methodology used.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10065302
MDR Text Key222559748
Report Number1823260-2020-01241
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot Number432844
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age23 YR
-
-