• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Model Number 02.124.417
Device Problem Break (1069)
Patient Problems Pain (1994); Non-union Bone Fracture (2369); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Additional device product codes: hrs, hwc.Reporter is attorney.Complainant part is not expected to be returned for manufacturer review / investigation.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that initially on (b)(6) 2018, the patient underwent an open reduction and internal fixation of the left femur due to a comminuted fracture of the midshaft.The patient has a boot on the right leg for another orthopedic issue in which the velcro of the boot got stuck on the recliner handle causing her to fall twisting her left leg.The patient had pain and inability to ambulate afterward and was brought to the emergency department.The fracture was stabilized using (1) 4.5mm variable angle- locking compression plate curved condylar plate 16 holes, left, 336mm, (6) synthes 1.7mm cerclage cables and (12) synthes screws.She tolerated the procedure well, was awakened, transferred to the recovery room in good condition.The patient was then taken to a skilled nursing facility after discharge where she spent 2 and half months completing inpatient rehab.Afterward, she was released to home where she then spent another month working with a home health physical therapist (2 times per week) and later was asked to progress to outpatient therapy.She has been doing well in her recovery with increasing mobility and with no pain in the left thigh.Radiographs performed on (b)(6) 2019, demonstrated the hardware in a good position and fracture well aligned.On (b)(6) 2019, the patient sought to consult due to pain and difficulty walking.The pain was described as sharp, aching, soreness continuously aggravated by certain movements like walking and weight-bearing.The pain score was 10/10.3 days prior to the consultation, the patient jammed her leg into a glass top coffee table.She did not think much of it then the following day, she was doing exercises and had severe pain when she tried to walk afterward.X-rays on the same day demonstrated a failure of hardware with a broken plate with all of the screws proximally and distally and all of the cables intact.On (b)(6) 2019, the patient underwent removal of one (1) 4.5mm variable angle- locking compression plate curved condylar plate 16 holes, left, 336mm, six (6) 1.7mm cerclage cables and twelve (12) screws and revision open reduction and internal fixation of midshaft left femur due to hardware failure and nonunion.During the operation, the fracture and hardware noted to be markedly comminuted and displaced.The majority of the fracture seemed to have healed but the transverse fracture had not healed.The patient tolerated the procedure well, was awakened, and transferred to the recovery room in good condition.Concomitant devices reported: 5.0mm variable angle locking screw 70mm (part number: 02.231.270, lot number: unknown, quantity: 3) 5.0mm variable angle locking screw 46mm (part number: 02.231.246, lot number: unknown, quantity: 1) 1.7mm cocr cable with ti crimp 750mm-sterile (part number: 611.105.01s, lot number: p307701, quantity: 2) 1.7mm cocr cable with ti crimp 750mm-sterile (part number: 611.105.01s, lot number: p177539, quantity: 2) 1.7mm cocr cable with ti crimp 750mm-sterile (part number: 611.105.01s, lot number: p177263, quantity: 1) 5.0mm variable angle locking screw 42mm (part number: 02.231.242, lot number: unknown, quantity: 3) 1.7mm cable with crimp 750mm-sterile (part number: 298.801.01s, lot number: unknown, quantity: 2) 5.0mm variable angle locking screw 32mm (part number: 02.231.232, lot number: unknown, quantity: 1) 5.0mm variable angle locking screw 40mm (part number: 02.231.240, lot number: unknown, quantity: 1) 5.0mm variable angle lockng screw 75mm (part number: 02.231.275, lot number: unknown, quantity: 2) unknown, quantity: 1).This report is for one (1) 4.5mm va-lcp curved condylar plate/16 hole/336mm/left.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.B7 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b7; d6a; d6b.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT
Type of Device
IMPLANT,FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10066277
MDR Text Key191985839
Report Number2939274-2020-02377
Device Sequence Number1
Product Code JDP
UDI-Device Identifier10886982042958
UDI-Public(01)10886982042958
Combination Product (y/n)N
PMA/PMN Number
K110354
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number02.124.417
Device Catalogue Number02.124.417
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/30/2020
Initial Date FDA Received05/18/2020
Supplement Dates Manufacturer Received06/02/2021
07/16/2021
Supplement Dates FDA Received06/04/2021
08/13/2021
Patient Sequence Number1
Treatment
1.7MM CABLE WITH CRIMP 750MM-STERILE; 1.7MM CABLE WITH CRIMP 750MM-STERILE; 1.7MM COCR CABLE WITH TI CRIMP 750MM-STERILE; 1.7MM COCR CABLE WITH TI CRIMP 750MM-STERILE; 1.7MM COCR CABLE WITH TI CRIMP 750MM-STERILE; 1.7MM COCR CABLE WITH TI CRIMP 750MM-STERILE; 1.7MM COCR CABLE WITH TI CRIMP 750MM-STERILE; 5.0MM VARIABLE ANGLE LOCKNG SCREW 32MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 40MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 42MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 42MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 42MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 46MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 70MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 70MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 70MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 75MM; 5.0MM VARIABLE ANGLE LOCKNG SCREW 75MM
Patient Outcome(s) Required Intervention;
Patient Weight60
-
-