Model Number 397002-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Power Problem (3010)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce 5287 initial.
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that during the daily system check, the companion 2 driver did not recognize the wall power and did not recognize the two external batteries that were installed.The external batteries showed full charge from the fuel gauge and the driver was plugged into the ac outlet.
|
|
Manufacturer Narrative
|
The customer-reported issue of the driver having an error for not detecting external batteries or external power (wall power) was not able to be reproduced during investigation testing.This investigation found that the driver functioned as intended and passed all requirements, which includes tests to confirm that the driver display shows a plug icon (indicating connection to external main a/c power) and for the hospital cart and driver to both display a plug icon indicating connection to a/c power when the driver is docked to the hospital cart.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.The external batteries used by the customer at the time of the reported event were not returned and therefore could not be evaluated as part of this investigation.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5287 follow-up report 1.
|
|
Search Alerts/Recalls
|