It was reported that during the set up of an knee arthroscopy with meniscal repair procedure, the shafts of both loop-ended stylets were detached from the proximal metal ends.Procedure was completed with the same device.No significant delay and no patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10 h3, h6: the reported disposable meniscus mender ii kit, used in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.An exact root cause cannot be determined with confidence without the return of the device.However, a trend of this nature has been observed with this product in the field, prompting a root cause investigation.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the complaint records was performed for the reported device, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|