Model Number 10-85-00
Device Problem Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 04/16/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.The pump was turned on and an error code different from the reported one was displayed.The error message experienced by the technician on site was associated to the was not controlled by the monitoring functions.The pump was then powered off and back on and no errors appeared.The technician in charge will replace the motor control board on a second visit.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that at the start up a s5 double head pump was giving an error message associated to a wrong direction of the pump.The user turned off the pump and did not use it.There was no patient involvement.
 
Manufacturer Narrative
H10: through follow-up communication livanova learned that the motor control board and the motor power amplifier will be replaced to solve the reported issue.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key10079420
MDR Text Key194049230
Report Number9611109-2020-00314
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Date Manufacturer Received06/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-