Model Number FGS-0313 |
Device Problems
Loss of or Failure to Bond (1068); Positioning Failure (1158)
|
Patient Problem
Aspiration/Inhalation (1725)
|
Event Date 05/05/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, they had two capsules which failed to attach and they fell off in the patient's airway, but they were able to retrieve them.They did not attempt to use a third capsule.
|
|
Manufacturer Narrative
|
Additional information: b3, d10, g4, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for evaluation.Visual inspection identified that only part of the device was returned.The delivery device was returned but not the capsule.No visible damages were noted and the plunger on the delivery device was fully released.Functional testing found that the device functioned normally and within specifications.It was reported that the capsule failed to attach to the patient's esophagus.A related device issue could not be confirmed.The most likely root cause could not be identified because no related problem was detected with the device.The evaluation found no potentially contributing factors, and the sample met all related specifications.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that the product has met all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, they had two capsules which failed to attach and they fell off in the patient's airway, but they were able to retrieve them.They did not attempt to use a third capsule.There was nothing unusual about the patient or the procedure, and an endoscopy had been performed prior to the procedure that showed the esophagus to be normal.The delivery system and capsule will be returned for investigation.
|
|
Manufacturer Narrative
|
Additional information: b5, d8, e3, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|