Model Number 0165SI24
Device Problems Inaccurate Flow Rate (1249); Defective Component (2292); No Flow (2991)
Patient Problem Urinary Tract Infection (2120)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the patient has been hospitalize four times for urinary infection while using the 2 way 5cc foley catheter.It was stated that the doctor suggested the catheters are defective and will not drain.Patient received antibiotics for the uti per doctors visit.
 
Manufacturer Narrative
The reported event was inconclusive.As no sample was returned for evaluation.However, the potential root cause for this failure mode could be user related (example: salt accumulation)/ block drainage lumen/ no drainage eye.The device history record was reviewed, and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "sterile: unless package is opened or damaged.Do not use if package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Visually inspect the product for any imperfections or surface deterioration prior to use.Valve type: use luer slip syringe.Do not use needle.Recommended inflation capacities: 3cc balloon: use 5cc sterile water, 5cc balloon: use 10cc sterile water, 30cc balloon: use 35cc sterile water.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back, and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol".Correction: b2.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Event Description
It was reported, that the patient has been hospitalized four times.For urinary infection, while using the 2 way 5cc foley catheter.It was stated, that the doctor suggested the catheters are defective, and will not drain.Patient received antibiotics for the uti, per doctors visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10079894
MDR Text Key191729230
Report Number1018233-2020-03302
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016790
UDI-Public(01)00801741016790
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 07/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number0165SI24
Device Catalogue Number0165SI24
Device Lot NumberNGCV4288
Was Device Available for Evaluation? No
Date Manufacturer Received06/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-