There was no death or device malfunction associated with the defibrillation event.Device evaluation summary: monitor sn (b)(4) and electrode belt sn (b)(4) were returned to the distributor for evaluation.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the monitor and ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the audio messaging and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.The pulse delivery circuitry test verified proper delivery of full energy 150j biphasic pulse.The functional testing confirmed proper ecg acquisition and pulse delivery functionality.There is no indication of a product malfunction.Manufacture date: monitor - (b)(4) - 06/27/2017, belt - (b)(4) - 03/28/2018.
|