• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US CORDCUTTER *EA; SUTURE CUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US CORDCUTTER *EA; SUTURE CUTTER Back to Search Results
Model Number 214646
Device Problem Dull, Blunt (2407)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/12/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the sales rep via email that during an unknown procedure the cordcutter will not cut.The cutting mechanism will not engage.Another device was used to complete the procedure.No patient consequences or surgical delay reported.The device is available for evaluation.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary - according to the information provided, it was reported that during an unknown procedure the cordcutter will not cut.The cutting mechanism will not engage.The complaint device is not being returned, therefore unavailable for a physical evaluation.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device [13n01] number, and no non-conformances were identified. at this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SUTURE CUTTER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10081094
MDR Text Key205926807
Report Number1221934-2020-01436
Device Sequence Number1
Product Code FZT
UDI-Device Identifier10886705004270
UDI-Public10886705004270
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214646
Device Catalogue Number214646
Device Lot Number13N01
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/01/2020
Patient Sequence Number1
-
-