U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET Back to Search Results
Catalog Number 66800039
Device Problems Difficult to Insert (1316); Failure to Power Up (1476)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2020
Event Type  Injury  
Event Description
It was reported that the versajet ii console is not powering on.No procedure was being performed when this was found hence no patient was involved.
 
Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Event Description
It was reported that prior to an incision drainage the versajet ii console was tested but was not powering on.Two hand-pieces were tested but neither of them fitted securely, so the device did not work.Since this occurred right before the procedure, the doctor had to use a simpulse irrigator instead for washing out the wound.
 
Manufacturer Narrative
The device, which was used in a procedure, was not returned for evaluation.A relationship between the device and the reported incident could not be confirmed, therefore the root cause is undetermined at this time, if the product is returned in the future the complaint can be reopened and evaluated.However factors that are known to contribute to the events reported, this may include loose connection issues to the electrical outlet.The ifu has been reviewed and contains comprehensive guides on the operation, use and inspection, including steps to be taken in the event of no intermittent electrical power, the associated risk file contains harms related to versajet, however as no harm was alleged within this complaint a review is not required.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.Our quality department will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10084838
MDR Text Key191972625
Report Number8043484-2020-00528
Device Sequence Number1
Product Code FQH
UDI-Device Identifier04582111153524
UDI-Public4582111153524
Combination Product (y/n)N
PMA/PMN Number
K110958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800039
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date04/27/2020
Date Manufacturer Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-