Philips has received through the food and drug administration medwatch program a voluntary event report submitted by chandler regional medical center (report number mw (b)(4)).In this report, it was reported: cath lab (cardiac catheterization lab) would not boot up at start of the day.Failure of the ippc (image processing pc).This device has failed or.12/12/19 and 2/5/20 as well.Manufacturer: can only replace the part, which then fails, and then several months later it fails again.This time there was not a patient on the table yet.Philips has started an investigation for this complaint.
|
Philips has investigated this complaint.Philips has inspected the system on site and it was identified that the ip-pc (image processing pc) failed.Analysis of the log files has also confirmed the failure of the ip pc.The defective ip-pc was replaced and the system was returned to use in good working order.Based on trending analysis there is no negative trend in the replacement rate for the ip-pc.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|