PHILIPS HEALTHCARE ALLURA XPER FD20; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 722012
Device Problem Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still ongoing for this event.When the investigation is completed a follow-up will be sent to the fda.
 
Event Description
Philips has received through the food and drug administration medwatch program a voluntary event report submitted by chandler regional medical center (report number mw (b)(4)).In this report, it was reported: cath lab (cardiac catheterization lab) would not boot up at start of the day.Failure of the ippc (image processing pc).This device has failed or.12/12/19 and 2/5/20 as well.Manufacturer: can only replace the part, which then fails, and then several months later it fails again.This time there was not a patient on the table yet.Philips has started an investigation for this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.Philips has inspected the system on site and it was identified that the ip-pc (image processing pc) failed.Analysis of the log files has also confirmed the failure of the ip pc.The defective ip-pc was replaced and the system was returned to use in good working order.Based on trending analysis there is no negative trend in the replacement rate for the ip-pc.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD20
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
MDR Report Key10086826
MDR Text Key192465732
Report Number3003768277-2020-00028
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
PMA/PMN Number
K130638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722012
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-