• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 7N8377
Device Problem Failure to Disconnect (2541)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported 2 one-link non-dehp y-type standard bore catheter extension sets were unable to disconnect.The issue was further described as "the product does not come apart" and it appears the "one-link is seized to the y-connector".This issue was identified during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, a video of the sample was provided for evaluation.Visual inspection was performed to the video using the naked eye which observed the difficulty of remove the luer activated device from the winged female.The reported condition was verified.The cause of the condition could no be determined.Due to the nature of the returned sample, no additional testing could be performed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10088017
MDR Text Key192042001
Report Number1416980-2020-02902
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412091808
UDI-Public(01)00085412091808
Combination Product (y/n)N
PMA/PMN Number
K132734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number7N8377
Device Lot NumberDR19L11044
Was Device Available for Evaluation? No
Date Manufacturer Received06/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number8040700¿12/04/19¿001-C
Patient Sequence Number1
-
-