• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. SPINAL ANES/EPIDURAL CATH KIT; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. SPINAL ANES/EPIDURAL CATH KIT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Catalog Number AK-05560
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the catheter tip was left back in the patient.The patient had to return to surgery to retrieve foreign body.
 
Event Description
It was reported that the catheter tip was left back in the patient.The patient had to return to surgery to retrieve foreign body.
 
Manufacturer Narrative
Qn#(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.Visual inspection could not be performed as no sample was returned by the customer for investigation.The customer did provide a photo of a lidstock for the reported kit.The ifu for this kit, e-17019-109a; rev.07, was reviewed as a part of this complaint investigation.The ifu warns the end user, "never advance catheter more than 5 cm beyond the needle tip.Advancing catheter more than 5 cm increases the likelihood of catheter-related complications." the ifu warns the user, "the ifu warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal." a corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.The customer did provide a photo of a lidstock for the reported kit.A device history record review was performed on the epidural catheter with no evidence to indicate a manufacturing related issue.Therefore, the potential cause of the catheter breaking could not be determined based upon the information provided and without a sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINAL ANES/EPIDURAL CATH KIT
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10098806
MDR Text Key193022306
Report Number1036844-2020-00177
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue NumberAK-05560
Device Lot Number23F20B0085
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/21/2020
Initial Date FDA Received05/28/2020
Supplement Dates Manufacturer Received07/09/2020
Supplement Dates FDA Received07/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-