Model Number CI-1400-01 |
Device Problem
Device Slipped (1584)
|
Patient Problem
No Information (3190)
|
Event Date 07/18/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing intermittency, static sounds, and non-auditory sensations, followed by loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
Additional information: section b.3, d.7, & d.10 the recipient's device was reportedly explanted.The recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient was reportedly reimplanted with another advanced bionics cochlear device.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.It is believed that the failure of the implantable cochlear stimulator device was caused by a loss of hermetic seal, which was concluded from the residual gas analysis data.This ultimately caused the device to cease functioning.This version of the hires device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|
|
|