ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problem Device Slipped (1584)
Patient Problem No Information (3190)
Event Date 07/18/2020
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing intermittency, static sounds, and non-auditory sensations, followed by loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
 
Manufacturer Narrative
Additional information: section b.3, d.7, & d.10 the recipient's device was reportedly explanted.The recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient was reportedly reimplanted with another advanced bionics cochlear device.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.It is believed that the failure of the implantable cochlear stimulator device was caused by a loss of hermetic seal, which was concluded from the residual gas analysis data.This ultimately caused the device to cease functioning.This version of the hires device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10099006
MDR Text Key193292173
Report Number3006556115-2020-00397
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/01/2007
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/28/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/11/2020
11/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age16 YR
-
-