ZIMMER BIOMET, INC. UNKNOWN LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Naturally Worn (2988)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Implant date: 2004.Concomitant medical products: item #: unknown, unknown head, lot #: unknown.Item #: unknown, unknown cup, lot #: unknown.Item #: unknown, unknown stem, lot #: unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The product is still implanted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2020-01727.0001822565-2020-01931.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.X-ray review identified constraining ring fracture, liner wear, and osteolysis at the tip of the femoral stem.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent an initial right hip surgery approximately 16 years ago.The patient has been indicated for a revision due to instability and liner wear with liner constraining ring fracture.The surgeon plans to revise patient to a constrained cup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key10106163
MDR Text Key195130968
Report Number0001822565-2020-01932
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/29/2020
Initial Date FDA Received06/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-