MAUDE Adverse Event Report: CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET,ADMINISTRATION,INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET,ADMINISTRATION,INTRAVASCULAR Back to Search Results
Model Number 2420-0007
Device Problems Fluid/Blood Leak (1250); Stretched (1601)
Patient Problems Venipuncture (2129); No Consequences Or Impact To Patient (2199)
Event Date 04/27/2020
Event Type  Injury  
Manufacturer Narrative
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.(b)(4).
 
Event Description
It was reported from the icu that after cleaning up the patient, the clinician went to restart the continuous renal replacement therapy dialysis machine.It was then noted that the pre-filter tubing for sodium bicarbonate 400ml solution (infusion rate set at 150ml over 4 hours) separated at the upper fitment and led to blood return and leak of fluids, which "hit the ceiling and floor".The charge nurse came in for assistance and replaced the tubing.However, the same issue occurred upon loading the new set on the same pump module.A third tubing was loaded onto the pump and it began forming a balloon at the pump segment.The rn was splashed in the face with blood and iv fluid when the tubing separated.The nursing supervisor was notified, blood specimen for (b)(6) and (b)(6) were sent the laboratory.The clinician was then notified that the (b)(6) and (b)(6) tests were both (b)(6).The clinician was further instructed to self-monitor at home for the next two weeks for any signs and symptoms of changes in temperature, cough, and shortness of breath and to notify the supervisor if the clinician had any of these symptoms.There were no adverse effect caused to the patient from this event.The rn returned to work without restrictions.
 
Manufacturer Narrative
As no samples were received for evaluation, no investigation was able to be performed.The root cause of two sets¿ tubing separated at the upper fitment and leaked fluids was not identified.The root cause of the third set formed a balloon at the pump segment was not identified.The device history record for primary set model: 2420-0007, lot: 20033185 was performed.The search showed a total of (b)(4) units in 1 lot were built on 19 june 2019.There were no related qn¿s (quality notifications) issued during the production build of this set for the failure mode reported for the given time frame.The device history record for primary set model: 2420-0007, lot: 20033186 was performed.The search showed a total of (b)(4) units in 1 lot were built on 19 june 2019.There were no related qn¿s (quality notifications) issued during the production build of this set for the failure mode reported for the given time frame.
 
Event Description
It was reported from the icu that after cleaning up the patient, the clinician went to restart the continuous renal replacement therapy dialysis machine.It was then noted that the pre-filter tubing for sodium bicarbonate 400ml solution (infusion rate set at 150ml over 4 hours) separated at the upper fitment and led to blood return and leak of fluids, which "hit the ceiling and floor".The charge nurse came in for assistance and replaced the tubing.However, the same issue occurred upon loading the new set on the same pump module.A third tubing was loaded onto the pump and it began forming a balloon at the pump segment.The rn was splashed in the face with blood and iv fluid when the tubing separated.The nursing supervisor was notified, blood specimen for human immunodeficiency virus (hiv) and hepatitis were sent the laboratory.The clinician was then notified that the hiv and hepatitis tests were both negative.The clinician was further instructed to self-monitor at home for the next two weeks for any signs and symptoms of changes in temperature, cough, and shortness of breath and to notify the supervisor if the clinician had any of these symptoms.There were no adverse effect caused to the patient from this event.The rn returned to work without restrictions.
 
Manufacturer Narrative
The device history record for primary set model 2420-0007 lot 20033186 was performed.The search showed a total of (b)(4) in 1 lot were built on (b)(6) 2020.There were no related qn¿s (quality notifications) issued during the production build of this set for the failure mode reported for the given time frame.
 
Event Description
It was reported from the icu that after cleaning up the patient, the clinician went to restart the continuous renal replacement therapy dialysis machine.It was then noted that the pre-filter tubing for sodium bicarbonate 400ml solution (infusion rate set at 150ml over 4 hours) separated at the upper fitment and led to blood return and leak of fluids, which "hit the ceiling and floor".The charge nurse came in for assistance and replaced the tubing.However, the same issue occurred upon loading the new set on the same pump module.A third tubing was loaded onto the pump and it began forming a balloon at the pump segment.The rn was splashed in the face with blood and iv fluid when the tubing separated.The nursing supervisor was notified, blood specimen for human immunodeficiency virus (hiv) and hepatitis were sent the laboratory.The clinician was then notified that the hiv and hepatitis tests were both negative.The clinician was further instructed to self-monitor at home for the next two weeks for any signs and symptoms of changes in temperature, cough, and shortness of breath and to notify the supervisor if the clinician had any of these symptoms.There were no adverse effect caused to the patient from this event.The rn returned to work without restrictions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Type of Device
SET,ADMINISTRATION,INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key10107343
MDR Text Key194689697
Report Number9616066-2020-01731
Device Sequence Number1
Product Code FPA
UDI-Device Identifier07613203021012
UDI-Public7613203021012
Combination Product (y/n)N
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/11/2023
Device Model Number2420-0007
Device Catalogue Number2420-0007
Device Lot Number20033186
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/01/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/14/2020
08/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
8015,8100, TD (B)(6) 2020.
Patient Outcome(s) Other;
-
-