US SURGICAL PUERTO RICO PREMIUM SURGICLIP; CLIP, IMPLANTABLE Back to Search Results
Model Number 134044
Device Problem Mechanical Problem (1384)
Patient Problem No Information (3190)
Event Date 05/18/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, there was a clip malformation.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for evaluation.The instrument was received partially applied with one remaining clip.Visual inspection of the instrument revealed no abnormalities.Functionally, the instrument was fired and proper clip formation was confirmed.The jaw and handle moved smoothly through the firing cycle and returned to the open position.The clip also loaded properly in the jaw and the interlock feature engaged appropriately.It was reported that there was clip malformation.A device related issue could not be confirmed.The most likely root cause could not be established because no issue was detected.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that products meet all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM SURGICLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key10114344
MDR Text Key193635375
Report Number2647580-2020-01726
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057531
UDI-Public10884521057531
Combination Product (y/n)N
PMA/PMN Number
K143644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number134044
Device Catalogue Number134044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2020
Date Manufacturer Received06/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-