• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE AORTIC VALVE; TISSUE, HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE AORTIC VALVE; TISSUE, HEART-VALVE Back to Search Results
Model Number 8300AB
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Complete Heart Block (2627)
Event Date 09/01/2019
Event Type  Injury  
Manufacturer Narrative
Udi (b)(4).The device was not returned to edwards for evaluation as it remains implanted.The device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Peri-procedural arrhythmias, heart block, and other conduction disturbances are common in patients with underlying cardiovascular disease and/or conduction abnormalities.They can be exacerbated with standard perioperative medications, anesthesia, and/or instrumentation of the heart.Temporary pacemakers are inserted in all patients undergoing aortic valve replacement (avr).It is not uncommon for patients to have short term/reversible periods of heart block or arrhythmias following the procedure while their heart recovers from cardiopulmonary bypass.In many cases, the temporary pm is left in the patient for a short time following the procedure and then subsequently removed prior to discharge.Following surgical aortic valve replacement (avr), new-onset bundle branch block has been reported in 16% to 32% of patients and the need for permanent pacemakers in 3% to 8% of patients.The reason for post-operative av block after surgical avr is related to injury to the cardiac conduction system during surgical excision of the adjacent diseased valve and annular tissue.The close anatomical relationship between the aortic valvular complex and the branching atrioventricular bundle explains the possible development of conduction abnormalities following prosthetic aortic valve procedures.A manufacturing related issue was not identified.A definitive root cause could not be determined.If additional information is received a supplemental mdr will be submitted.Edwards lifesciences will continue to monitor all reported events.No further actions are required at this time.
 
Event Description
Edwards received information that the third degree atrioventricular block was detected after six (6) days form the implantation surgery of this 27mm valve.This device was originally implanted to correct aortic stenosis due to severe calcification.After an implant duration of approximately three (3) months, pacemaker was implanted to the patient.The patient status was reported as ¿recovering¿.The device was not returned for evaluation as it remained implanted.Multiple cardiac surgical procedure: coronary artery bypass grafting at implant.The surgeon commented that this event was device related as this event occurred after implantation surgery.
 
Manufacturer Narrative
Corrected data- updated h6 (result).
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE AORTIC VALVE
Type of Device
TISSUE, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key10121441
MDR Text Key194109408
Report Number2015691-2020-12068
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/23/2021
Device Model Number8300AB
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age72 YR
Patient Weight72
-
-