MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT TRACHE DIRECT INTERFACE; BZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT TRACHE DIRECT INTERFACE; BZA Back to Search Results
Model Number OPT870
Device Problem Gas/Air Leak (2946)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
Ps339988.The opt870 adult tracheostomy interface is used to deliver humidified oxygen to patients via tracheostomy.The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's tracheostomy tube.Method: the complaint opt870 interface was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photo provided by the customer, and our knowledge of the product.Results: visual inspection of the provided revealed a hole in the opt870 tube.Conclusion: without the complaint device, we were unable to conclusively determine the cause of the reported failure.However, the damage was most likely caused by tubing being punctured with a blunt object.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and deformation.Any product that fails the visual inspection is rejected.The subject device would have met the required specification at the time of production.The user instructions accompanying opt870 adult tracheostomy interface state the following: - "to ensure loading on tracheostomy tube is kept to a minimum, make sure lanyard is fitted appropriately." - "do not crush or stretch tube.".
 
Event Description
A distributor in (b)(6) reported on behalf of a healthcare facility, via a fisher & paykel healthcare (f&p) field representative, that the tubing of a opt870 adult tracheostomy interface was leaking air during use.It was also reported that there were some holes on the tube.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT TRACHE DIRECT INTERFACE
Type of Device
BZA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, ca 
4534000142
MDR Report Key10121850
MDR Text Key199488247
Report Number9611451-2020-00515
Device Sequence Number1
Product Code BZA
UDI-Device Identifier09420012432148
UDI-Public010942001243214810
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT870
Device Catalogue NumberOPT870
Device Lot Number2100749259
Was Device Available for Evaluation? No
Date Manufacturer Received05/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-