• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 0DEG 34MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 0DEG 34MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC Back to Search Results
Catalog Number NLS-340000B
Device Problem Insufficient Information (3190)
Patient Problems Inflammation (1932); Pain (1994); Reaction (2414)
Event Date 07/26/2013
Event Type  Injury  
Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported elevated levels of cobalt and chromium is considered to be under the scope of this recall.No further investigation is required.An event regarding pain involving a rejuvenate modular device was reported.The event was confirmed.Method & results: device evaluation and results: device evaluation was not performed as no devices were received.Device history review: review of device history records found the devices in the reported lot were accepted into final stock with no reported discrepancies.Complaint history review: the complaint databases show there have been other events for the reported lot.Similar events have occurred for the catalog number and product family.These events were determined to be associated with ra 2012-067.Conclusions: voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported pain is considered to be under the scope of this recall.No further investigation is required.Not available.
 
Event Description
It was reported that the patient has upper side and hip pain.Additional information received from legal on 5/12/2020: the plaintiff alleges that the rejuvenate implanted on (b)(6) 2009 failed due to elevated levels of cobalt and chromium revealed in bloodwork.The revision has not been scheduled yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LRG TAP PRI MOD NCK 0DEG 34MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
IE   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key10125055
MDR Text Key194315499
Report Number0002249697-2020-01039
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2013
Device Catalogue NumberNLS-340000B
Device Lot Number25520802
Was Device Available for Evaluation? No
Date Manufacturer Received05/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2090-2012
Patient Sequence Number1
Patient Outcome(s) Other;
-
-