Model Number CD3369-40Q
Device Problem Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/20/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented for initial implant.During procedure, the physician attempted to implant the implantable cardioverter defibrillator, but the left ventricular lead would not fit into the header.After several failed attempts, the physician decided to use another device instead.Patient was stable post procedure.
 
Manufacturer Narrative
The reported field event of a header anomaly was confirmed.The header of the device was tested on the bench and epoxy leakage was found in the middle area of seals and contacts from the connector is4.The cause of the issue was due to a manufacturing anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10125849
MDR Text Key194215956
Report Number2017865-2020-07166
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000094928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/20/2020
Initial Date FDA Received06/05/2020
Supplement Dates Manufacturer Received12/15/2020
Supplement Dates FDA Received12/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age85 YR
-
-