• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Device Problem Material Deformation (2976)
Patient Problem Capsular Bag Tear (2639)
Event Date 01/09/2020
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during a cataract surgery with an intraocular lens (iol) implantation, the temporal haptic was noted to be above the iris and attempted to be rotated in.During manipulation of the lens, the anterior rent radialized and caused a large posterior tear.The iol was removed from the eye from the main wound after enlargement.The procedure was completed.A lens was inserted and anterior vitrectomy was performed.Additional information was provided indicating that in the surgeon's opinion the cause of the event was a combination of patient eye movement and a kinked haptic.The patient's issue have not resolved yet, but planning to have a sutured iol after covid.The patient will require repeat surgery to suture iol in eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key10125934
MDR Text Key194306392
Report Number1119421-2020-00834
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.220
Device Lot Number12732877
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2020
Initial Date FDA Received06/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age32 YR
-
-