MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-33J
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported that the user found a volume of water pooled in the corrugated tube and the water trap during use on a patient.Therefore, the adaptor was replaced with a new one.No patient harm was reported.
 
Manufacturer Narrative
(b)(4).One unit of 031-33j "nebulizer adaptor 033,sterile,japanese" was received for analysis.Signs of use were observed since the sample was not received on its original package.A visual exam was performed and no defects were observed.Functional testing was also performed and no issues were encountered.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.The unit functioned as intended.
 
Event Description
It was reported that the user found a volume of water pooled in the corrugated tube and the water trap during use on a patient.Therefore, the adaptor was replaced with a new one.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key10130159
MDR Text Key194794767
Report Number3004365956-2020-00114
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number031-33J
Device Lot Number74H1900010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Initial Date Manufacturer Received 05/21/2020
Initial Date FDA Received06/08/2020
Supplement Dates Manufacturer Received07/08/2020
Supplement Dates FDA Received07/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NONE REPORTED.; NONE REPORTED.
-
-