C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. FOLEY CATHETER; LATEX FOLEY CATHETER Back to Search Results
Catalog Number 1853SI20
Device Problems Inadequate Instructions for Healthcare Professional (1319); Device Markings/Labelling Problem (2911)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that there was a labeling error on the packaging of "dufour" probes (red latex foley catheter).External packaging stated dufour, labeling of the primary packaging indicated "delinotte".
 
Manufacturer Narrative
The reported event was confirmed as project execution error.The pouch print artwork was printed as ¿delinotte¿ instead of dufour.The error was unintentionally happened during the internal artwork file change at the packaging printing machine during udi eu project execution.The device history record was reviewed and found a possible manufacturing issue that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "bard urological catheter.Warning: do not use ointments or lubricants having a petroleum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml if sterile water.Or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon.Gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.".H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that there was a labeling error on the packaging of "dufour" probes (red latex foley catheter).External packaging stated dufour, labeling of the primary packaging indicated "delinotte".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. FOLEY CATHETER
Type of Device
LATEX FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10130805
MDR Text Key194771667
Report Number1018233-2020-03677
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 07/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1853SI20
Device Lot NumberMYDX0082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2020
Date Manufacturer Received06/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-