ST. JUDE MEDICAL, INC. ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100014514
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/21/2020
Event Type  Injury  
Manufacturer Narrative
One ensite velocity¿ system amplifier was returned for investigation.Power was applied to the returned amplifier which successfully completed the post (power on self-test) and the system status light changed to green.System log files were reviewed which revealed both voltage out of range and temperature out of range errors for the catheter amplifier board in the slot thirteen location.Review of the system log files and current board status revealed the catheter amplifier board in the slot 13 location was not active with multiple faults stored.Based on the information provided to abbott and the investigation performed, the root cause of the self-test issue and subsequent procedure cancellation was isolated to abnormal functionality of the catheter amplifier pca in the slot thirteen location.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.
 
Event Description
During the procedure, the amplifier would not boot and displayed an amber flashing light.Multiple reboots were attempted with no resolution and the procedure was cancelled.Multiple power receptacles were attempted and the air filter was cleaned but the light remained amber.There were no adverse consequences to the patient due to the cancellation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key10132835
MDR Text Key194472525
Report Number2184149-2020-00076
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 06/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number100014514
Device Catalogue Number100014514
Device Lot Number7025155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/21/2020
Initial Date FDA Received06/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-