This supplemental report is being submitted to provide the review of the device history records (dhr).Please see updated sections: g4, g7, h2, h3,h4,h6 and h10.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Per device evaluation, the user reported issue of "blurry image was confirmed.The failure was due to chipped cover glass, fiber breakage, excessive debris underneath cover glass on distal tip and loose meniscus.Olympus will continue to monitor complaints for this device.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct the serial number of the device.Based on the results of the investigation, the physical damage found on the device was most likely due to user mishandling.To avoid inadvertent damage, the device's instructions for use advises: "study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.".
|